Utrasound as a method for confirming the positioning of enteral catheters in critically ill patients
DOI:
https://doi.org/10.11152/mu-4392Keywords:
enteral nutrition, ultrasound, gastrointestinal intubation, nasoenteral catheterAbstract
Aim: To evaluate the agreement between the bedside ultrasound in a single epigastric window and the plain X-ray to confirm the positioning of the enteral catheter in critically ill patients.
Material and methods: This was an observational, cross-sectional study conducted in two Intensive Care Units of a university hospital. The ultrasound exams were carried out immediately after the introduction of the enteral catheter, using only the epigastric window, with an injection of 5 ml of air associated with 5 ml of saline solution. In all cases, the plain radiography was taken to confirm the positioning of the enteral catheter and to define the beginning of nutritional therapy.
Results: This study included 83 patients, the positioning of the enteral catheter was confirmed by plain radiography in all cases and by ultrasound in 81 (97.6%) patients. The median duration of the ultrasound exam was 2 (2-3) minutes, while the time spent between the request for the X-ray and the release of the exam for a doctor’s appointment was 225 (120-330) minutes.
Conclusion: Bedside ultrasound proved to be an effective, quick, and safe method to confirm the position of the enteral catheter in critically ill patients.
Downloads
Published
Issue
Section
License
All rights are reserved to the Medical Publishing House of the „Iuliu Hatieganu" University of Medicine and Pharmacy, Cluj-Napoca; no parts of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise without the permission of the editor.