Prospective evaluation of Quasistatic Ultrasound Elastography (USE) compared with Baseline US for parotid gland lesions: preliminary results of elasticity contrast index (ECI) evaluation

Authors

  • Vito Cantisani Department of Radiology, Oncology, and Anatomy Pathology, Policlinico Umberto I, University Sapienza, Rome, Italy
  • Emanuele David Department of Biomedical Sciences and Morphological and Functional Imaging, University of Messina, Policlinico “G. Martino”, Messina, Italy
  • Armando De Virgilio Department of Sensory Organs, ENT Section, “Sapienza” University of Rome, Rome, Italy
  • Paul S. Sidhu Department of Radiology, King’s College Hospital, Denmark Hill, London, UK
  • Hektor Grazhdani Associazione dei Cavalieri Italiani Sovrano Militare Ordine di Malta; Poliambulatorio Roma Eur, Rome, Italy
  • Antonio Greco Department of Sensory Organs, ENT Section, “Sapienza” University of Rome, Rome, Italy
  • Marco De Vincentiis Department of Sensory Organs, ENT Section, “Sapienza” University of Rome, Rome, Italy
  • Alessandro Corsi Department of Molecular Medicine, Policlinico Umberto I, University Sapienza, Rome, Italy
  • Flaminia De Cristofaro Department of Radiology, Oncology, and Anatomy Pathology, Policlinico Umberto I, University Sapienza, Rome, Italy
  • Luca Brunese Department of Medicine and Health Sciences, University of Molise, Campobasso,Italy
  • Fabrizio Calliada University Hospital of Pavia, Pavia, Italy
  • Alfredo Blandino Department of Biomedical Sciences and Morphological and Functional Imaging, University of Messina, Policlinico “G. Martino”, Messina, Italy
  • Mario Tombolini Department of Sensory Organs, ENT Section, “Sapienza” University of Rome, Rome, Italy
  • Giorgio Ascenti Department of Biomedical Sciences and Morphological and Functional Imaging, University of Messina, Policlinico “G. Martino”, Messina, Italy
  • Roberto Stramare University of Padova , Department of Medicine, Padova, Italy
  • Marcello Caratozzolo Department of Radiology, Oncology, and Anatomy Pathology, Policlinico Umberto I, University Sapienza, Rome, Italy
  • Ferdinando D’Ambrosio Department of Radiology, Oncology, and Anatomy Pathology, Policlinico Umberto I, University Sapienza, Rome, Italy

DOI:

https://doi.org/10.11152/mu-923

Keywords:

parotid, ultrasound, elastography, parotidectomy, elastosonography

Abstract

Aims: To assess the diagnostic value of elasticity contrast index (ECI) elastography, in comparison with conventional ultrasound for the differentiation of benign and malignant parotid lesions. Material and methods: Sixty-three consecutive patients with parotid gland tumors, treated surgically at a single tertiary center were enrolled. Ultrasound evaluation consisted of B-mode, color-Doppler and quasistatic ultrasound elastography (USE), conducted with the ECI technique. For each lesion the echogenicity, margins, vascularization, and capsulation were considered to determine benign or malignant lesions. Histology was considered the gold standard. Results: Using conventional parameters, the observer suggested malignant lesions in 20 cases and benign lesions in 43 cases (accuracy: 61.8%). ECI>3.5 alone was the most accurate parameter (accuracy: 90.5%), with sensitivity of 93.7% and specificity of 89.4%. However, when combined with the other US criteria, no statistically significant diagnostic accuracy improvement was achieved. Conclusions: Our results show that USE with ECI index measurement can help to discriminate preoperatively benign from malignant lesions, with the exception of pleomorphic adenoma which is stiff. We suggest that USE can be used as an additional tool to conventional US evaluation of salivary gland lesions.