(258 KB, PDF file)
Aim: Reliable liver stiffness measurements (RLSM) using Transient Elastography (TE) with the standard M probe are dif- cult to obtain in overweight (BMI≥25kg/m2) and obese (BMI>30kg/m2) patients. The aim of our paper was to assess the fea- sibility of TE in daily practice using both M and XL probes. Material and method: We studied retrospectively 3235 patients with chronic liver disease assessed by TE rst by the M probe (standard probe – transducer frequency 3.5 MHz), and if the measurements were unreliable, with the XL probe (transducer frequency 2.5 MHz). Reliable measurements were de ned as the median of 10 valid measurements with a success rate ≥ 60% and an interquartile range < 30%. Results of liver elasticity were expressed in kiloPascals (kPa). Results: RLSM by M probe were obtained in 62.2% (2015/3235) patients, and by XL probe in 1011/1220 (80%) of patients with unreliable measurements by M probe; thus we obtained RLSM in 93.5% of 3235 cases. In overweight patients we obtained RLSM in 89.9% (1039/1156) cases: in 63.1% (729) by M probe and in 26.8% (310) by XL probe. In obese patients we obtained RLSM in 83.8% (746/890): in 18.4% (164) by M probe and in 65.4% (582) by XL probe. Thus, by using both probes, RLSM were obtained in 1785 (87.2%) of overweight and obese patients. Conclusion: The feasibility of the M probe was 62.2% in our Department. Reliable measurements using M or XL probe allowed the evaluation of liver stiffness in 93.5% of cases. By using both M and XL probes, reliable LSM by TE can be obtained in the majority of obese and overweight patients (87.2%).